Correction of target-controlled infusion following wrong selection of emulsion concentrations of propofol
Korean Journal of Anesthesiology 2014³â 66±Ç 5È£ p.377 ~ p.382
äÀ±Á¤(Chae Yun-Jeong) - Ajou University School of Medicine Department of Anesthesiology and Pain Medicine
Á¶Çѹü(Joe Han-Bum) - Ajou University School of Medicine Department of Anesthesiology and Pain Medicine
ÀÌ¿øÀÏ(Lee Won-Il) - Ajou University School of Medicine Department of Anesthesiology and Pain Medicine
±èÁø¾Æ(Kim Jin-A) - Ajou University School of Medicine Department of Anesthesiology and Pain Medicine
¹Î»ó±â(Min Sang-Kee) - Ajou University School of Medicine Department of Anesthesiology and Pain Medicine
Abstract
Background: We investigated the correction methods following wrong-settings of emulsion concentrations of propofol as a countermeasure against erroneous target-controlled infusions (TCI).
Methods: TCIs were started with targeting 4.0 ¥ìg/ml of effect-site concentration (Ceff) of propofol, and the emulsion concentrations were selected for 2.0% instead of 1.0% (FALSE1-2, n = 24), or 1.0% instead of 2.0% (FALSE2-1, n = 24). These wrong TCIs were corrected at 3 min after infusion start. During FALSE1-2, the deficit was filled up while injecting after equilibrium (n = 12), or while overriding (n = 12). During FALSE2-1, the overdose was evacuated while targeting Ceff (n = 12) or targeting plasma concentration (Cp) (n = 12). The gravimetrical measurements of TCI reproduced the Cp and Ceff using simulations. The reproduced Ceff at 3 min (Ceff-3min) and the time to be normalized within ¡¾ 5% of target Ceff (T¡¾5%), were compared between the correction methods.
Results: During the wrong TCI, Ceff-3min was 1.98 ¡¾ 0.01 ¥ìg/ml in FALSE1-2, and 7.99 ¡¾ 0.05 ¥ìg/ml in FALSE2-1. In FALSE1-2, T¡¾5% was significantly shorter when corrected while overriding (3.9 ¡¾ 0.25 min), than corrected after equilibrium (6.9 ¡¾ 0.05 min) (P < 0.001). In FALSE2-1, T¡¾5% was significantly shorter during targeting Cp (3.6 ¡¾ 0.04 min) than targeting Ceff (6.7 ¡¾ 0.15 min) (P < 0.001).
Conclusions: The correction methods, based on the pharmacokinetic and pharmacodynamic characteristics, could effectively and rapidly normalize the wrong TCI following erroneously selections of the emulsion concentration of propofol.
Ű¿öµå
Drug delivery systems, Infusion pumps, Intravenous infusion, Propofol