Duodenum identification mechanism for capsule endoscopy.

Woo, Sang Hyo; Mohy-Ud-Din, Zia; Cho, Jin Ho
IEEE transactions on bio-medical engineering
2011Apr ; 58 ( 4 ) :905-12.
ÀúÀÚ »ó¼¼Á¤º¸
Woo, Sang Hyo -
Mohy-Ud-Din, Zia -
Cho, Jin Ho -
ABSTRACT
The aim of this study is to implement a duodenum identification mechanism for capsule endoscopes because commercially available capsule endoscopes sometimes present a false negative diagnosis of the duodenum. One reason for the false negative diagnosis is that the duodenum is the fastest moving part within the gastrointestinal tract and the current frame rate of the capsule is not fast enough. When the capsule can automatically identify that it is in the duodenum, the frame rate of the capsule can be temporarily increased to reduce the possibility of a false negative diagnosis. This study proposes a mechanism to identify the duodenum using capacitive proximity sensors that can distinguish the surrounding tissue and transmit data using RF communication. The implemented capsule (D11 mm × L22 mm) was smaller than the commercially available capsule endoscopes, and power consumption was as low as 0.642 mW. Preexperiments were conducted to select an appropriate electrode width in order to increase the signal-to-noise ratio (SNR), and in vitro experiments were conducted to verify whether the implemented capsule could identify the duodenum within 3 s. The experiment showed that the identification rate of duodenum was 93% when the velocity of the capsule was less than 1 cm/s.
Capacitive proximity sensor, capsule endoscope, duodenum, telemetry, electrostatic
MESH
*Algorithms, Animals, Capsule Endoscopy/*instrumentation, Duodenum/*anatomy & histology/*physiology, Electric Capacitance, *Electrodes, Equipment Design, Equipment Failure Analysis, Humans, In Vitro Techniques, Plethysmography, Impedance/*instrumentation, Reproducibility of Results, Sensitivity and Specificity, Swine
¸µÅ©

ÁÖÁ¦ÄÚµå
ÁÖÁ¦¸í(Target field)
¿¬±¸´ë»ó(Population)
¿¬±¸Âü¿©(Sample size)
´ë»ó¼ºº°(Gender)
Áúº´Æ¯¼º(Condition Category)
¿¬±¸È¯°æ(Setting)
¿¬±¸¼³°è(Study Design)
¿¬±¸±â°£(Period)
ÁßÀç¹æ¹ý(Intervention Type)
ÁßÀç¸íĪ(Intervention Name)
Å°¿öµå(Keyword)
À¯È¿¼º°á°ú(Recomendation)
Identification rate of duodenum was 93% when the velocity of the capsule was less than 1 cm/s.